Enhancing Battery Performance: Utilizing FPI ICP-OES to Accurately Assess Metal Impurities in Electrolytes
Pub Date:2024-01-24 Source: View Num:753



The electrolyte is crucial for lithium-ion batteries, affecting their temperature, energy capacity, cycle efficiency, and safety. Lithium hexafluorophosphate (LiPF6) is a core material used in electrolyte production. However, LiPF6 electrolyte may contain impurities like hydrogen fluoride, water, and metal ions that can negatively impact battery performance. Excessive metal impurity ions can decrease the battery's reversible capacity and prevent proper electrode passivation, potentially leading to battery damage. Therefore, strict limits are set for metal element content in LiPF6 electrolyte.


To accurately measure metal element content, conventional testing methods involve dilution, which may underestimate actual concentrations and introduce new impurities. However, the standard method (HG/T 4067-2015) recommends direct measurement without digestion. Our research center follows this method by diluting LiPF6 electrolyte with 20% ethanol and using inductively coupled plasma optical emission spectrometry (ICP-OES) with cooling and auto-sampling systems. This approach provides reliable analysis for LiPF6 electrolyte samples with good reproducibility and recovery rates.


Experimental part

Instrument

Table 1 Inductively Coupled Plasma Spectrometer

Model: ICP-OES EXPEC-6500

Configuration: EXPEC 6500D Organic Sampling System



Table 2 Detection Parameters of Inductively Coupled Plasma Spectrometer

Parameters

Setpoint

RF Power

1150

Atomized Gas Flow

0.6

Auxiliary Gas Flow

1

Cooling Gas Flow

14

Flush/Analyze Pump Speed

(rpm)

50

Response Time(s)

Smart Points

Observation Method

Axial

TEC Refrigeration Temperature (°C)

-2


Reagents and Standards

Reagents: Electronic grade anhydrous ethanol

Purified water: 18.2 MΩ·cm deionized water

Standard solutions: Al, Ca, Co, Cr, Cu, Fe, Hg, Mg, Mo, Ni, Na, Pb, S, Zn single-element standard solutions, 1000 ug/mL, provided by the National Nonferrous Metals Research Institute.


Sample Preparation
Weigh 2.0g of lithium hexafluorophosphate electrolyte and dilute it with 20% ethanol to a total weight of 10.0g.


Standard Curve and Detection Limit

Select appropriate analytical spectral lines for the target elements and plot the standard curve. The test results showed a linear correlation coefficient greater than 0.9990 for the target elements. The concentration corresponding to three times the standard deviation of the measured values from the blank samples, obtained by continuous analysis of 11 replicates, is considered the detection limit of the instrument. The linear correlation coefficient, analytical spectral lines, and detection limits for each target element are shown in Table 3.


Table 3: Analytical Spectral Lines, Linear Correlation Coefficients, and Detection Limits for the Target Elements

Elements

Analytical Lines (nm)

Linear correlation coefficient r

Method Limit of Detection (mg/kg)

AI

167.079

0.99999

0.038

Ca

317.933

0.99997

0.048

Co

228.616

0.99994

0.036

Cr

206.55

0.99991

0.034

Cu

224.7

0.99991

0.063

Fe

259.94

0.99999

0.044

Hg

194.227

0.99996

0.096

Mg

280.27

0.9999

0.049

Mo

281.615

0.99991

0.046

Ni

221.647

0.99995

0.039

Na

589,592

0.99998

0.069

Pb

220.353

0.99999

0.096

S

182.624

0.99994

0.107

Zn

213.856

0.99984

0.023


Precision Testing Method

Seven replicate samples of lithium hexafluorophosphate electrolyte, after being spiked with standards, were subjected to secondary testing. The results indicated that the relative standard deviations (RSD) for all elements were less than 5.0%. This demonstrates the excellent precision of the method. The precision testing results for each element in the spiked lithium hexafluorophosphate electrolyte samples are presented in Table 4.


Table 4: Precision Testing Results of Spiked Lithium Hexafluorophosphate Electrolyte Samples (unit: mg/Kg)

Element

Parallel-1

Parallel-2

Parallel-3

Parallel-4

Parallel-5

Parallel-6

Parallel-7

Average

RSD/%

Al

0.904

0.924

0.937

0.897

0.914

0.922

0.911

0.916

1.46

Co

0.184

0.175

0.174

0.189

0.194

0.194

0.191

0.186

4.56

Cr

0.217

0.221

0.208

0.204

0.221

0.231

0.209

0.216

4.37

Cu

0.201

0.195

0.189

0.214

0.216

0.207

0.2

0.203

4.84

Fe

0.251

0.261

0.248

0.237

0.264

0.255

0.249

0.252

3.57

Mo

0.199

0.187

0.182

0.197

0.206

0.209

0.21

0.199

5.46

Na

0.991

0.979

0.99

0.988

0.991

0.98

0.995

0.988

0.607

Ni

0.224

0.218

0.21

0.225

0.234

0.231

0.217

0.223

3.75

S

1.24

1.27

1.19

1.17

1.23

1.25

1.31

1.24

3.81

Zn

0.207

0.198

0.21

0.205

0.215

0.204

0.197

0.205

3.1



Actual Sample Spike Recovery Testing

Spike recovery testing was conducted on two different samples of lithium hexafluorophosphate electrolyte. Each sample was spiked with appropriate concentrations based on the element content, as shown in Table 6. The spike recovery rates for each sample ranged from 90.7% to 108%.


Table 5: Sample Testing Results and Spike Recovery Rates (unit: mg/kg)

Sample

Element

Average

RSD/%

Spike Recovery: 1%

Hexagas lithium phosphate
electrolyte-1

Al

N.D

-

91.4

Co

N.D

-

93

Cr

N.D

-

105

Cu

N.D

-

102

Fe

0.22

1.39

104

Mo

N.D

-

97.5

Na

0.397

4.03

90.7

Ni

0.208

1.47

91.5

S

0.832

0.734

108

Zn

N.D

-

104

Hexagas lithium phosphate
electrolyte-2

Al

N.D

-

92.1

Ca

N.D

-

104

Cr

N.D

-

104

Cu

N.D

-

100

Fe

0.23

1.26

94.5

Hg

0.67

4.89

90.9

Mg

0.6

2.5

94

Na

0.45

4.06

107

Pb

N.D

-

103

Zn

N.D

-

105



Conclusion

In this experiment, a method was established to determine the content of 14 elements (lead, iron, copper, zinc, chromium, aluminum, sodium, calcium, magnesium, mercury, sulfur, cobalt, nickel, and molybdenum) in lithium hexafluorophosphate electrolyte, which was diluted with 20% ethanol and analyzed using ICP-OES. The experimental results showed that the linear correlation coefficients of the established calibration curves were all greater than 0.9990. The precision testing of the target elements in the actual samples showed relative standard deviations (RSD) below 5.0%. Additionally, the spike recovery rates for the target elements in the actual samples ranged from 90% to 108%. The detection limits for the elements ranged from 0.023 to 0.107 mg/kg. These results indicate that the precision and accuracy of the sample testing were satisfactory, and this method can be applied to determine the content of iron, sodium, magnesium, mercury, sulfur, nickel, molybdenum, and other elements in lithium hexafluorophosphate electrolyte samples.

Quote Now
  • Name*
  • Email*
  • PhoneNumber*
  • Country/State*
  • Message*
Captcha*
Can’t read? Try another
I agree that my data from this contact form will be collected and processed to answer my request
Copyright 2023 FPI Group All Rights Reserved.浙ICP备11039119号 Powered ByHANSUN
Request for Quotation
Do you have any questions or requests?
Fill this form to contact our experts.
Name *
Email *
Message
Submit
XML 地图
XML 地图