AQMS-450 Carbon Dioxide Analyzer (CO2)

Product Summary:
FPI AQMS-450 carbon dioxide (CO2) analyzer measures ambient CO2 concentration by employing Non-dispersive infrared (NDIR) with gas filter correlation (GFC) method technology to realize ppm level CO2 concentration measurement. 
Quote Now
Product overview Product features
Product overview

FPI AQMS-450 carbon dioxide (CO2) analyzer measures ambient CO2 concentration by employing Non-dispersive infrared (NDIR) with gas filter correlation (GFC) method technology to realize ppm level CO2 concentration measurement. 


Principle
Infrared energy emitted by light source is passed through gas chamber containing the air sample, and the quantitative absorption of energy by CO in the sample cell is measured by corresponding detector.

GFC
GFC (Gas filter correlation) technology is utilized to remove interference caused by moisture and other backgrounds.
Two gas filled chambers are mounted on a rotating disc, which pass through an IR beam alternately. The measure chamber is filled with nitrogen while the reference chamber is filled with high concentration CO. IR beam then passes through the sample gas cell. The difference in absorbance is measured and provides a direct output of the gas concentration.

Data storage and analysis
Stored data are easily retrievable through the serial or ethernet port via PC client software, allowing operators to perform predictive diagnostics and enhanced data analysis by tracking parameter trends.


Standard Range

0~2000ppm 

Стандартный диапазон
0~2000ppm

Zero Noise

≤0.1ppm 

Нулевой шум
≤0.1ppm

Span Noise

≤5ppm 

Пролетный шум
≤5ppm

Lower Detectable Limit 

<0.2ppm 

Минимальный 

предел обнаружения

<0.2ppm

Precision 

≤5ppm 

Точность
≤5ppm

Response Time 

≤90s (T90) 

Время реакции
≤90сек (T90)
Output    

4~20mA, 

RS232, RS485, Ethernet


Вывод
4~20mA,RS232, RS485, Ethernet

Product features
  • The integrated purging device which eliminates the interference of environmental CO2 to the instrument measurement.
  • The White cavity design and special anti-corrosion technology on the surface of optical devices to give better stability and longer lifetime.
  • Wide measurement range with low detection limit, which can meet the needs of CO2 monitoring in most application.
  • Adopting GFC NDIR technology, measuring range is up to 2000 ppm.
  • High response, high repeatability, high accuracy and simple operation.
  • Support classified operation, automatically storing the calibration data and alarm information, etc.
Quote Now
  • Name*
  • Email*
  • PhoneNumber*
  • Country/State*
  • Message*
Captcha*
Can’t read? Try another
I agree that my data from this contact form will be collected and processed to answer my request
Copyright 2023 FPI Group All Rights Reserved.浙ICP备11039119号 Powered ByHANSUN
Request for Quotation
Do you have any questions or requests?
Fill this form to contact our experts.
Name *
Email *
Message
Submit
XML 地图
XML 地图